We manufacture a broad assortment of Flame Photometer. Due to the certified material and technology used in making of these photometers, the photometers deliver quality of highest order. Providing a lasting life, these photometers are vital in testing blood and soil samples to determine Sodium, Potassium, Lithium and Calcium. We offer different models of photometers like Microprocessor based and Digital Flame Photometers. These photometers are one of the contemporary instruments with features like PC connectivity through RS232 serial port connection. With the single aspiration of sample, measuring of two elements can be done simultaneously. Our product range finds application in the medical clinical labs, soil testing labs, milk plants etc.
CLICK THE LINK GIVEN BELOW TO VIEW/DOWNLOAD OUR LEAFLETS.
www.4shared.com/dir/5DsrzlE1/sharing.html
VSI make Digital Flame Photometer, model : (VSI-FP1 & VSI-FP2) is an economical, reliable, rugged, most accurate and precise instrument suitable for measurement of Sodium (Na) and Potassium (K) with the filters mounted on the disc. Calcium and Lithium are optional.
Digital Clinical Flame Photometer Model:'VSI-FP2' is a dual channel Flame Photometer used for simultaneous observation of Na and K in blood serum (Plasma) Urine and other type of biological fluids.
Technical Specifications : VSI-FP1
Main Unit |
VSI-FP1 |
Range |
Na : 0-100ppm K : 0-100ppm Ca : 15-100ppm (Optional) K : 10-100ppm (Optional) |
Sensitivity |
Sodium (Na) : 5ppm Potassium (K) : 5ppm Calcium (Ca) : 10ppm. (Optional) Lithium (Li) : 10ppm. (Optional) |
Display |
2½ Digit LED Display |
Accuracy |
+-2% upto 40 ppm +-1 digit +-5% above 40 ppm +-1 digit |
Ignition System |
With External Igniter |
Measurement System |
Measures one element at a time |
Repeatability |
+-2 Counts |
Detector |
Photocell/Photodiode |
Filters |
Narrow band interference glass filters |
Nebulizer |
Black bakelite or S. S., axial flow type |
Flame System |
LPG & Dry Oil free Air |
Combustion Gas |
LPG Controlled by precision needle type regulator |
Warm-up Time |
10 minutes |
Power Supply |
230 V +-10%, 50 Hz AC, Built-in-Voltage Stabilizer |
Compressor Unit |
VSI-FP1 |
Air Supply |
By Oil Free Mini Compressor Unit with pressure regulator |
Power |
230V +-10%, 50 Hz AC |
STANDARD ACCESSORIES |
Instrument with Na & K Filters, 5 Nos. Sample Holders, 1 No. Chimney Cover, 2 Nos. Tubes for Gas & Air Connection with 4 Nos. Hose Clamps and Instruction Manual. |
Technical Specifications : VSI-FP2
Main Unit |
VSI-FP2 |
Range |
Sodium (Na) : Serum 0-200 meq./l, Urine 0-200 meq./l Potassium (K) : Serum 0-10 meq./l Urine 0-100 meq./l Lithium (Lj) : Serum 0-2 meq./l (Optional) Calcium (Ca) : Urine 0-10 meq./l (Optional) |
Display |
Dual 2½ Digit LED Display |
Accuracy |
+-2% upto 100 meq. +-5% above 100 meq. |
Ignition System |
With External Igniter |
Measurement System |
Simultaneous measurement of 2 Elements on Dual LED Display |
Repeatability |
+-2 Counts |
Detector |
Photocell/Photodiode |
Filters |
Narrow band interference glass filters |
Nebulizer |
Black bakelite or S.S., axial flow type |
Flame System |
LPG & Dry Oil free Air |
Combustion Gas |
LPG controlled by precision needle type regulator |
Warm-up Time |
10 minutes |
Power Supply |
230 V +-10%, 50 Hz AC, Built-in-Voltage Stabilizer |
Compressor Unit |
VSI-FP2 |
Air Supply |
By Oil Free Mini Compressor Unit with pressure regulator |
Power |
230V +-10%, 50 Hz AC |
STANDARD ACCESSORIES |
Instrument with Na & K Filters, 5 Nos. Sample Holders, 1 No. Chimney Cover, 2 Nos. Tubes for Gas & Air Connection with 4 Nos. Hose Clamps and Instruction Manual. |
Microprocessor Flame Photometer (Model : VSI-601)
Microprocessor Flame Photometer, (Model : VSI-601) is an ideal instrument for the determination of Sodium (Na), Potassium (K), Calcium (Ca) and Lithium (Li). The instrument is supplied with two filters (Na & K) only (Ca & Li are optional – available at extra cost). It uses the latest Micro-controller technology and advanced engineering techniques so as to give enhanced accuracy and reproducibility. It has soft touch membrane key for ease of operations. The solution is aspirated through an automizer, air, sample and the fuel are mixed in the mixing chamber which is then sprayed as a very fine mist into the flame. The color of the flame is changed depending upon the concentration of elements present. Radiations from the flame passes through the sensing system and specific narrow band interference filter which permits only the characteristic radiation to pass to the photo-detector. The output of the photo-detector is then processed by the micro-controller and the final results are displayed on the digital display.
FEATURES
TECHNICAL SPECIFICATIONS : VSI-601
RANGE
General Range Serum Range
ppm meq/l meq/l
Na : 0-100 Upto 250meq/l, 1:100 dil. 0-200 meq/l, 1:100 dil.
K : 0-100 Upto 250meq/l, 1:100 dil. 0-10 meq/l, 1:100 dil.
Ca : 0-100 Upto 250meq/l, 1:100 dil. ----
Li : 0-100 Upto 250meq/l, 1:100 dil. 0-2 meq/l, 1:5 dil.
Urine Range (meq/l)
Na : 0-250 meq/l, 1:100 dil.
K : 0-100 meq/l, 1:100 dil.
Ca : 0-10 meq/l, 1:2 dil.
Li : ----
SENSITIVITY General Mode : Na: 2 ppm, K: 1 ppm,
Li: 1 ppm, Ca: 30 ppm
FILTERS Na & K Supplied
(Ca & Li are optional at extra cost)
RESOLUTION 0.1 ppm / meq.
REPRODUCIBILITY ±2% FS ± 2 Digits
CURVE FIR ACCURACY ±2% of FS
DISPLAY 20 x 4 Character alphanumeric
LCD display with backlit
AVERAGE TIME In-built in Software
FLAME SYSTEM LPG & Dry Oil Free Air
DETECTOR Photo diode
CALIBRATION Up to 5 Point Calibration with
Curve Fitting Software
GAS CONTROL Adjustable with knob
IGNITION SYSTEM Auto Ignition
AUTOMISER Axial Flow Type
AIR SUPPLY By oil free mini compressor unit
with pressure regulator
COMBUSTION GAS LPG Controlled by precision regulator
PRINTER Provision for attachment of Printer
RS 232 INTERFACE Available (PC software at extra cost)
SIZE(LxBxH) Main Unit : 365 x 245 x 220 mm Approx.
WEIGHT Main Unit : 7.5 Kg. (Approx.)
Compressor : 8 Kg. (Approx.)
POWER SUPPLY 230 V AC ± 10%, 50 Hz
Submit a request to get commercial offer by visiting our Contact Us page.
VSI ELECTRONICS PVT. LTD.
info.vsielectronicsmohali@gmail.com
F-330, PHASE VIII-B, SECTOR 74, INDUSTRIAL AREA, S.A.S. NAGAR, MOHALI - 160071, CHANDIGARH (INDIA).
+91-172-4627238,
+91-9855076463, 9814016463,
+91-9041931993